E Instruments International 2013 Test & Measurement Catalog

 

Embed or link this publication

Description

E Instruments Test & Measurement Catalog

Popular Pages


p. 1

e instruments international 2013 test measurement catalog www.e-inst.com f 215-750-1399 p 215-750-1212 langhorne pa 19047 a 402 middletown blvd ste 216

[close]

p. 2

calibration calibration calibration manometers mp 105 ­ mp 120 functions pressure selection of units manual automatic calibration hold function minimum and maximum values adjustable automatic shut-off adjustable backlight adjustable climatic parameters mp120 built-in calculation for velocity mp120 technical features measuring element overpressure allowed piezoresistif sensor mp 120 manometer air velocity mp105 1.4bar mp 120 250 mbar pressure connectors mp 120 Ø 6.2 mm barbed connectors made of nickelled brass mp 105 et 112 Ø 4.6 mm threaded connectors made of nickelled brass dimensions mm mp 105 ­ high range manometer display 156.2 147.7 housing keypad conformity power supply operating temperature storage temperature auto shut-off weight languages 2 lines lcd technology sizes 50 x 34.9 mm 1 line of 5 digits with 7 segments value 1 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection metal-coated with 5 keys electromagnetical compatibility nf en 61326-1 guideline 1 alkaline battery 9v 6lr61 from 0 to 50°c from -20 to +80°c adjustable from 0 to 120 min 6.7 oz english french top view 34.7 front view 50 50 side view 34.9 70.6 70.6

[close]

p. 3

specifications pressure mp 105 measuring units mesure kpa inwg mbar mmhg psi plages de exactitudes résolutions measuring range accuracy resolutions from 0 to ±200 inh2o 500 mbar ±0.5 of reading ±0.5 mbar 0.01 inh2o pressure air velocity pitot tube mp 120 kpa inwg mbar mmhg psi m/s fpm km/h from 0 to ±4 inh2o 1000 pa from 2 to 5 m/s from 5 to 40 m/s ±0.5 of reading ±2 pa ±0.7 m/s ±0.5 of reading ±0.3 m/s 0.01 inh2o 0.1 m/s all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation working principle piezoresistif sensor piezoresistif sensor is a diaphragm formed on a silicon substrate which bends with applied pressure and pressure generates millivoltage or millicurrent proportional to the pressure applied voltage pitot tube dynamic pressure is measured by pitot tube pd total pressure ­ static pressure velocity is calculated according to bernoulli simplified formula formula with temperature correction static pressure silicon layer total pressure pressure sensor vacuum vm/s k x 574,2 156842,77x p0 p en pa po barometric pressure in pa temperature in °c k pitot tube coefficient supplied with description pressure sensor from 0 to ±1000 pa pressure sensor from 0 to ±1000 mmh2o 465 mm 215 mm ftang ­ mp100 ­ 120 12/08 e ­ we reserve the right to modify the characteristics of our products without notice mp 105 mp 120 pressure sensor from 0 to ±500 mbar pressure sensor from 0 to ±2000 mbar pitot tube Ø 6mm length 300 mm 2x1 m clear tube Ø 4 x 6 mm 2x1 m silicone tube Ø 4 x 7 mm stainless steel tip Ø 6 x 100 mm calibration certificate transport case included optional accessories see related datasheet ce 100 protective cover with magnet and holding system j.t.c or j.y.c straight connections in t or y for tube Ø 5x8mm see related datasheet pitot tube available in many lengths Ø 3/6 or 8mm with or without temperature compensation warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department required for calibration repair p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com

[close]

p. 4

calibration ask about vane probe thermo-anemometers calibration certificate lv 101 107 110 functions air velocity and ambient temperature selection of units hold function minimum and maximum values automatic average airflow calculation airflow calculation with cone adjustable automatic shut-off adjustable backlight detection of flow direction automatic cfm measurements technical features measuring elements vane probe anemometer lv110 hall effect sensor ambient temperature pt100 class a 2 lines lcd technology sizes 50 x 34.9 mm 1 line of 5 digits with 7 segments value 1 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection metal-coated with 5 keys retractable length 450 mm up to 2.4 m when released electromagnetical compatibility nf en 61326-1 guideline 1 alkaline battery 9v included from 0 to 122°f from -4 to +176°f from -4 to +176°f adjustable from 0 to 120 min 6.7 oz english french display lv 101 ­ vane probe Ø 14 mm ­ fixed probe 331,7 mm 331.7 housing keypad cable conformity power supply operating temperature instrument operating temperature probe lv 107 ­ vane probe Ø 70 mm ­ fixed probe 12 inches 14 mm storage temperature auto shut-off weight languages 5.5 inches lv 110 vane probe Ø 100 mm ­ fixed probe 12 inches 5.5 inches

[close]

p. 5

specifications measuring units air velocity lv 101 Ø 14 mm lv 107 Ø 70 mm lv 110 Ø 100 mm measuring ranges from 0.8 to 3 m/s from 3.1 to 40 m/s from 0.3 to 3 m/s from 3.1 to 35 m/s from 0.25 to 3 m/s from 3.1 to 35 m/s accuracy ±3 of reading ±0.1m/s ±1 of reading ±0.3m/s ±3 of reading ±0.1m/s ±1 of reading ±0.3m/s ±3 of reading ±0.1m/s ±1 of reading ±0.3m/s resolutions 0.1 m/s 0.1 m/s 0.01 m/s 0.1 m/s 1 m3/h m/s fpm km/h m/s fpm km/h m/s fpm km/h air flow all models m3/h cfm l/s m3/h from 0 to 99999 m3/h ±3 of reading ±0.03 areacm2 ambient temperature all models °c °f from -20 to +80°c ±0.4 of reading ±0.3°c 0.1°c all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation working principle anemometer hall effect sensor rotation of the shaft of the vane powers a circular magnet of 8 poles a dual hall effect sensor placed next to the magnet senses the signals of magnetic field polarity transition the sensor signal is converted to electrical frequency and is proportional to the air velocity signal chronology allows the detection of rotation direction vane magnet with magnetic field in rotation vane in rotation supplied with 18 inches supplied with description optional lv 101 8.5 inches airflow lv 107 lv 110 polarity sensor 1 polarity sensor 2 hall effect sensor frequency vane velocity thermometer pt100 probe pt100 is a resistance with a positive temperature coefficient which varies according to the temperature the higher the temperature is the more the value of the resistance increases ie for 0°c 100 for 100°c 138,5 Ø 14 mm vane probe Ø 70 mm vane probe Ø 100 mm vane probe air flow cone k25 air flow cone k35-75-120 or 150 calibration certificate transport case accessories see related datasheet rts ce 100 protective cover with magnet and holding system k 25 air flow cone see related datasheet dimensions top view inches 2.7 70.6 telescopic extension for probe 1m long and bent at 90° rd 300 to be clipped extension for Ø14 mm vane rd300 Ø 10 mm length 300 mm 12 inches front view 2 50 1.36 side view 1.37 warranty period 5.8 instruments have 1-year guarantee for any manufacturing defect return to our service department for repair and/or calibration p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com

[close]

p. 6

lv120 lv130 vane probe thermo-anemometer ask about calibration calibration certificate functions air velocity and ambient temperature selection of units hold function minimum and maximum values automatic average airflow calculation airflow calculation with cone adjustable automatic shut-off adjustable backlight detection of flow direction technical features measuring elements display vane anemometer hall effect sensor ambient temperature pt100 class a 4 lines lcd technology sizes 50 x 34.9 mm 2 line of 5 digits with 7 segments value 2 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection metal-coated with 5 keys electromagnetical compatibility nf en 61326-1 guideline 1 alcaline battery 9v 6lr61 from 0 to 50°c from -20 to +80°c adjustable from 0 to 120 min 300 g french english two 2 models vane probe rotation +180° -90° increments of 90 ° lv 130 Ø 100 mm moving vane probe ­ fixed and rotating housing keypad conformity power supply operating temperature instrument lv 130 storage temperature auto shut-off weight languages side view lv 120 Ø 100 mm vane probe ­ fixed vane head

[close]

p. 7

specifications measuring units air velocity measuring ranges from 0.25 to 3 m/s from 3.1 to 35 m/s from 0 to 99999 m3/h accuracy ±3 of reading ±0.1m/s ±1 of reading ±0.3m/s ±3 of reading ±0.03 areacm2 resolutions 0.01 m/s 0.1 m/s 1 m3/h m/s fpm km/h air flow m3/h cfm l/s m3/h ambient temperature °c °f from 0 to +50°c ±0.4 of reading ±0.3°c 0.1°c all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation working principle anemometer hall effect sensor rotation of the shaft of the vane powers a circular magnet of 8 poles a dual hall effect sensor placed next to the magnet senses the signals of magnetic field polarity transition the sensor signal is converted to electrical frequency and is proportional to the air velocity signal chronology allows the detection of rotation direction vane magnet with magnetic field in rotation vane in rotation supplied with supplied with description Ø 100 mm vane probe moving Ø 100 mm vane probe air flow cone k25 k85 calibration certificate transport case polarity sensor 1 polarity sensor 2 option lv 120 lv 130 airflow hall effect sensor frequency vane velocity accessories see related datasheet k 25 ­ k 85 air flow cone see related datasheet thermometer pt100 probe pt100 is a resistance with a positive temperature coefficient which varies according to the temperature the higher the temperature is the more the value of the resistance increases ie for 0°c 100 for 100°c 138.5 ce 100 protective cover with magnet and holding system dimensions mm top view warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department for repair/calibration front view 106.1 side view p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com ftang ­ lv120-lv130 ­ 29/03/10 ­ rcs 24 périgueux 349 282 095 non-contractual document ­ we reserve the right to modify the characteristics of our products without prior notice 46.6 271.1

[close]

p. 8

calibration calibration ask about calibration certificate hot-wire thermo-anemometer vt 100 functions air velocity and ambient temperature selection of units hold function minimum and maximum values automatic average air flow calculation adjustable automatic shut-off adjustable backlight technical features measuring element hotwire air velocity thermistance with a negative temperature coefficient ambient temperature pt100 1/3 din 2 lines lcd technology sizes 50 x 34.9 mm 1 line of 5 digits with 7 segments value 1 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection metal-coated with 5 keys straight lg 2 m electromagnetical compatibility nf en 61326-1 guideline 1 alcaline battery 9v 6lr61 from 0 to 50°c from -20 to +80°c adjustable from 0 to 120 min 6.7 oz english french vt 100 ­ standard hot-wire 12 inches dimensions mm 8 mm display 147.7 housing keypad cable conformity power supply operating temperature storage temperature auto shut-off weight languages top view 34.7 front view 50 side view 50 34.9 70,6 70.6

[close]

p. 9

specifications measuring units air velocity hotwire measuring ranges 0,1 m/s accuracy ±3 of reading ±0.05m/s ±3 of reading ±0.2m/s resolutions m/s fpm km/h from 0.15 to 3 m/s from 3.1 to 30 m/s 0.01 m/s 0.1 m/s air flow m3/h cfm l/s m3/h ambient temperature °c °f from 0 to 99999 m3/h ±3 of reading ±0.03areacm2 1 m3/h from -20 to +80°c ±0.4 of reading ±0.3°c 0.1°c all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation working principle hotwire anemometer a wire is continuously heated at a superior temperature than ambient and continuously cooled by airflow constant temperature is maintained by a regulation circuit the heating current is proportional to the airflow velocity supplied with 465 mm included airflow hotwire regulation circuit heating current generator optional 215 mm description hotwire probe Ø 8mm length 300 mm airflow cone calibration certificate transport case vt 100 ftang ­ vt 100 ­ 05/08 b ­ we reserve the right to modify the characteristics of our products without notice i wire temperature current velocity regulation airflow display accessories see related datasheet ce 100 protective cover with magnet and holding system k 35 75 120 150 air flow cones see related datasheet thermometer pt100 probe pt100 is a resistance with a positive temperature coefficient which varies according to the temperature the higher the temperature is the more the value of the resistance increases ie for 0°c 100 for 100°c 138,5 bnf hotwire cleaning spray rd 300 hotwire extension to be clipped straight Ø 10 mm length 300 mm warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com

[close]

p. 10

calibration calibration thermo-hygrometers hd 100 hd 150 functions relative humidity dew point and ambient temperature selection of units temperature and dew point hold function minimum and maximum values adjustable automatic shut-off large ice blue adjustable backlight technical features measuring element display hygrometry and temperature cmos sensor 2 lines lcd technology size 50 x 34.9 mm 1 line of 5 digits with 7 segments value 1 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection or ip65 with food industry protective cover metal-coated with 5 keys retractable length 18 inches 450 mm up to 8 ft 2.4 m when released electromagnetical compatibility nf en 61326-1 guideline 1 alkaline battery 9v 6lr61 included from 0 to 122°f from -4 to +158°f from -4 to +176°f adjustable from 0 to 120 min 6.7 oz english french housing keypad cable conformity power supply operating temperature instrument operating temperature probe hd 100 ­ hygrometry probe ­ fixed probe black 10 5.5 0.5 storage temperature auto shut-off weight languages hd 150 ­ hygrometry probe ­ fixed probe white 10 5.5 0.5

[close]

p. 11

specifications measuring units relative humidity rh measuring range accuracy resolutions 0.1 rh from 5 to 95 rh accuracy repeatability linearity hysteresis ±1.8%rh from 15°c to 25°c factory calibration incertainty ±0.88 rh temperature dependence ±0.04 x t-20 %rh if t<15°c or t>25°c dew point °c td °f td ambient temperature from -40 to +158°ftd ±0.8 of reading ±0.6°ftd 0.1°ftd °c °f from -40 to +158°f ±0.4 of reading ±0.3°f 0.1°f all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation as per nfx 15-113 and the charter 2000/2001 hygrometers gal guaranteed accuracy limit which has been calculated with a coverage factor value of 2 is ±2.88%rh between 18 and 28°c on the measuring range from 5 to 95%rh sensor drift is less than 1%rh/year working principle capacitive sensing element for relative humidity measurement inside the probes a capacitive polymer layer reacts with the humidity present between two metal layers which cover a glass substract water absorption is a function of relative humidity of the surrounding environment and modifies the dielectric constant the measured signal is directly proportionnal to the relative humidity and is dependent on the atmospheric pressure supplied with ftang ­ hd 100 ­ hd150 ­ 01/08 a ­ we reserve the right to modify the characteristics of our products without notice included optional description 18.3 8.5 hd 100 hd 150 porous metal electrode humidity sensitive polymer layer electrode connection hygrometry probe Ø 13 mm protective white cover for food industry ip65 soft vinyl transport case capacity of relative humidity sensor relative dielectric permittivity humidity dependent void permittivity electrodes area optional hard abs case available main electrode glass substract electrodes spacing h relative humidity accessories see related datasheet ce 100 protective cover with magnet and holding system rts telescopic extension for probe 1m long and bent at 90° semiconductor temperature sensor the direct tension of a silicon diode is vbe vg01-t/t0 vbe0t/t0 nkt/qlnt0/t kt/qlnic/ic0 dimensions top view 70.6 70.6 warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department front view 50 34,7 side view 50 147.7 34.9 p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com

[close]

p. 12

hm100 supplied with fitting certificate pin moisture meter functions indication of humidity percentage for 4 types of different material hardwood softwood concrete and plaster bricks minimum and maximum values for each type of material protection cap and auto-test technical features display 4 lines lcd technology sizes 50 x 34.9 mm 2 line of 5 digits of 7 segments value 2 line of 5 digits of 16 segments unit shock-proof made of abs ip54 protection metal-coated with 5 keys electromagnetical compatibility nf en 61326-1 guideline 1 alcaline battery 9v 6lr61 neutral gas from 0 to 50°c from -20 to +80°c adjustable from 0 to 120 min 6.7 oz english french dimensions mm top view 34.7 162.7 162.7 housing keypad conformity power supply environment operating temperature storage temperature auto shut-off weight languages front view 70.6 70.6 22 side view 10.1

[close]

p. 13

specifications number 1 material softwood detail birch beech spruce larch cherry tree walnut oak pine maple ash pin douglas measuring range from 10.0 to 93.7 accuracy ±1 2 3 4 hardwood concrete and plaster bricks from 8.0 to 78.5 from 1.0 to 2.5 from 0.0 to 22.9 ±1 ±1 ±1 all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation working principle measurement of the electrical resistance the way how to determinate material moisture used today is the measurement of the wood electrical resistance a withered material is highly resistant to electrical current indeed water presence in material modifies the reaction of this material when it is submitted to an electrical current the more there is water in material the lower is its resistance and current circulation is easier if we consider this variation of resistance according to material moisture we can observe a precise regularity which apart from a few exceptions follows a similar curve with almost all wood species concrete plaster and bricks so a measurement instrument able to indicate material resistance can provide a direct measurement of present moisture with the help of a given scale supplied with ftang ­ hm100 ­ 29/03/10 ­ rcs 24 périgueux 349 282 095 non-contractual document ­ we reserve the right to modify the characteristics of our products without prior notice supplied with description protection cap two sets of pins fitting certificate transport case calibration certificate option hm 100 s working principle the principle consists on making pass an electrical current between two pins that are pushed in the wood as the figure below water is conductive so the more the wood will be wet the more the current will pass accessories see related datasheet ph100 one set of two additional pins ce 100 protective cover with magnet and holding system figure 1 measurement principle of electrical resistance warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department for calibration/repair p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com

[close]

p. 14

thermocouple thermometers ask about calibration certificate tk 100 ­ tk102 ­ tk150 tk152 functions temperature selection of units hold function simplified mode function minimum and maximum values adjustable automatic shut-off large adjustable ice blue backlight delta t tk 102 &152 only adjustable alarms auto-hold function technical features measuring elements display thermocouple k j or t class 1 2 lines lcd technology sizes 50 x 34.9 mm 1 line of 5 digits with 7 segments value 1 line of 5 digits with 16 segments unit shock-proof made of abs ip54 protection and ip65 with food industry protective cover metal-coated with 5 keys electromagnetical compatibility nf en 61326-1 guideline 1 alkaline battery 9v 6lr61 included from 0 to 122°f from -4 to +176°f adjustable from 0 to 120 min 6.7 oz english french special water proof protective cover for food market housing keypad conformity power supply operating temperature storage temperature auto shut-off weight languages tk 100 ­ 1 channel tk150 ­ white for food market tk 102 ­ 2 channels tk152 ­ white for food market +

[close]

p. 15

specifications measuring units thermocouple probes see related datasheet measuring ranges accuracy resolutions thermocouple k thermocouple j thermocouple t °c °f °c °f °c °f from -328 to 2372°f from -158 to 1382°f from -328 to 752°f ±1.1°f or ±0.4 of reading ±0.8°f or ±0.4 of reading ±0.5°f or ±0.4 of reading 0.1 °f 0.1 °f 0.1 °f all accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions or carried out with required compensation the accuracy is expressed either by a deviation in °c or by a percentage of the value concerned only the bigger value is considered working principle according to the seebeck effect when two wires composed of different metals are joined at both ends an electric circuit is formed the voltage increases with temperature i.e thermocouple k chromel hot junction alumel supplied with description thermocouple probe inputs waterproof protective cover for food industry ip65 calibration certificate soft vinyl transport case included tk 100 tk 102 optional tk 150 tk 152 dimensions mm large choice of temperature probes see related datasheet ambient food industry contact penetration penetration general use etc 18 8.5 top view 70.6 70.6 accessories see related datasheet 34.7 ce 100 protective cover with magnet and holding system bn see related datasheet black ball Ø 150mm with junction for temperature probe Ø 4.5mm further dimensions available front view 50 50 side view rts telescopic extension for probe 1m long and bent at 90° gst silicone heat-conductive grease for temperature probes warranty period instruments have 1-year guarantee for any manufacturing defect return to our service department 147.7 p 21 5-7 50 -1212 f 215 -750-1 39 9 a 402 middletown b lvd s te 216 la ng horne pa 19 047 e instruments international www.e-inst.com ftang ­ tk 100 tk 150 ­ 11/08 b ­ we reserve the right to modify the characteristics of our products without notice 34.9