p. 1
part 6 equipment instruments teaching aids required for college of pharmacy 1 student microscope routine laboratory microscope with phase contrast and dark field scope of supply · 1 x microscope stand with mechanical stage 75 x 30r · 2 x halogen bulb 6v 30w · 1 x binocular tube 35deg/20 · 1 x conversion filter · 1 x dust cover · 1 x abbe condenser 0.9/1.25 with h/d/ph 1-2-3 turret · 1 x ultra condenser · 1 x condenser carrier · 1 x eyepiece 10/20 br · 1 x eyepiece 10x/20 br foc · 2 x eyepiece eyecup · 1 x achroplan 10x/0.25 ph1 · 1 x achroplan 20x/0.45 ph2 · 1 x achroplan 40x/0.65 ph2 · 1 x achroplan 100x/1.25 oil ph3 · 1 x achroplan 100x/1.25 oil iris incubator · anaerobic workstation · large stainless steel condensate free work area · two adjustable shelves · multiple electrical outlets at rear of unit · fluorescent lighting · circulating fan · centrally located glove ports allow access to all areas · clear vinyl front · built in incubator 2.72 cu.ft with two shelves · operating temperature variable +4ºc to 70ºc ± 0.3ºc · anaerobosis by gas mixture with palladium catalyst · front opening exchange chamber with control panel showing chamber functions and status · auto sequence gas changeover to anaerobis by vacuum pump and triple purge · thermostat alarm · charcoal filter · dessicant wafer · gas and vacuum connections · requires anaerobic gas mixture 85%n2/10%h2/5 co2 high purity 30 2 1
[close]
p. 2
· nitrogen high purity 4 kymograph with assembly this assembly includes the universal kymograph with stimulator plus a comprehensive set of accessories with a 165 page applications manual with detailed instructions for over 30 experiments some of the possible experiments include frog heart in situ preparation · recording of cardiac activity · effect of stimulation of the heart refractory period compensatory pause · demonstration of conduction path stimulus ligatures · effects of stimulating the vagus nerve cardiac inhibition · effect of drugs muscle-nerve preparation · effect of stimulating voltage simple muscle twitch · effect of stimulation frequency tetanus · staircase phenomenon treppe · effect of repeated stimulation fatigue · effects of drugs on muscles · effect of pre and after-loading on muscle efficiency for locust work a special cork mounting block is supplied which wedges vertically in the muscle chamber metabolic cage this metabolic cage for rats enables collection of behavioral information blood urine and feces when part of the culex animal handling system or it can be used alone strictly for metabolism studies to collect urine and feces it is likewise an ideal cage for continuous infusion studies for toxicology or metabolism studies or both this is a replacement part or add-on for those wanting extra clean cages on hand to cycle through their experimental protocols the metabolic rat cage can be used on the basic metabolism stand offered to the right each cage is supplied as a complete kit a stand must be purchased separately the metabolic cage for rats comes complete with the following components metabolic rat cage stainless steel food dish rat feces fetcher stainless steel mesh basket pyrex rat urinal cup reusable chilling chamber for scintillation vials rat water tube stainless steel rat funnel metabolic rat cage lid food dispenser block off plate for rat metabolic cage 1 set 5 5 1
[close]
p. 3
6 myograph board myograph system with dual wire system normalization module illuminated chamber digitally controlled temperature temperature resolution efficacy 0.1oc ergonomically designed sliding wrist rest transducer safety lock probe sensor analog electrical output linearity is 0 to 100 internal frequency response is selectable from 1000hz to 10hz measure contraction/relaxation of vessels 50300 m oxygenation facility volume chamber 5-10 ml optical window easy fixing of wire two chamber system temperature probe data acquisition software inbuilt force transducer spiro meter measurements vc fvc fev1 fev1 pef pif f50 f25 fef25-75 i50 r50 fet mvvind [all expiratory measurements with baseline post-bronchodilator %predicted %change and normal range storage 90 complete tests including both pre and post-flow/volume and volume/time loops all fully addressable printer output 320 dot internal thermal printer predicted values knudson or crapo display high resolution graphic lcd 128x64 pixels transducer micro direct bi-directional digital volume resolution 10 ml volume 0.03l/s flow accuracy to ats recommendations standardization of spirometry 1994 update for flows and volumes meets all known current international standards for spirometry test nos vc-unlimited best reported fvc-maximum 9 all reported interface rs232 input/output power supply input 100 to 250v 50 to 60 hz output 12v 1.5a tissue bath assembly 2x reservoir 2x reservoir holder 2x transducer holder 2x coil holder 2x warming coil specify 5ml 10ml 20ml or 30ml size 2x bath holder 2x tissue holder glass 2x tissue holder stainless steel 4x triangle tissue holder stainless steel 4x tissue clip stainless steel 3 7 2 8 1
[close]
p. 4
2x 2x 2x 2x 2x 2x 9 dip cage bath specify 5ml 10ml 20ml or 30ml size oxygen filter micrometer assembly mounting accessory kit base station with support rods force transducer choose tsd125 model c/d/e orf 2 1 2 4 3 10 frog cage 11 mice cage 12 rabbit restrainer 13 sthethograph with mariestambor part 7 equipment required for national institute of liver and gastro intestinal diseases nilgid 1 fibro scan fibroscan for non invasive diagnosis of liver fibrosis with standard accessories and 4 yars extended comprehensive warranty 1 1 1 1 1 1 1 1 1 1 part 9
[close]
p. 5
equipment instruments required in microbiology diagnostic laboratory 1 vortex mixture · vortex mixer to operate when pressure is applied to the rubber cup which receives the vessel containing the materials to be mixed · variable speed control to accept vessels up to 25mm diameter 2 ph meter electronics · unit to have buffer sachets electrode arm electrode lead and shorting plug · available modes · ph mv absolute relative mv temperature timer 4 digit display 8 character alphanumeric prompting led display measurement of ph millivolts and temperature 3 shaker description orbital shakers are made out of powdercoated heavy gauge ms body instrument is designed in such a way that it can accommodate 50 ml 100 ml 250 ml 500 ml,1000 ml flasks d.c motor is used to get the accurate speed control and low noise technical specification motion orbital speed 30-250 rpm controller digital platform size 410 mm x 410 mm input voltage 230 v.50 hz a.c supply 4 anaerobic jar · unit to be manufactured from stainless steel · the seal between the jar and lid to incorporate a neoprene o ring the large lid clamp to locate into a dimple on the lid to allow rapid operation and minimum clamp pressure · the jar to be fitted with a side arm with a viewing device which gives positive secondary vacuum measurement 5 slide warmer · used to assist in the handling of sections and embedded tissue · can also be used for fixing bacterial smears · nominal temperature range 80 deg c 6 filtration unit with steel funnel sartorius stainless steel quality high-grade stainless steel b.s 304s31|aisi 304 dimensions in mm l|h|d 3-branch manifold 435|103|120 6-branch manifold 910|103|120 max operating pressure vacuum only 2 1 2 3 3 1
[close]
p. 6
sterilization by autoclaving max 134°c by dry heat max 180°c by flaming by other methods acc to iso 8199 materials lid funnel base part °© filter support clamp and tap made of stainless steel silicone flat gasket silicone lid seal flow rate per filter station for water at 90 vacuum 200 ml/min with 0.2 µm membrane filter 600 ml/min with 0.45 µm membrane filter filtration area 12.5 cm© suitable membrane filter diameter 50 mm 47 mm if using a 47 mm frit filter support 6980103 outlet spouts individual system 10 mm outside diameter inlets branches only female threads tr 20 x 2 outlet branches only hose nipple dn 10 stainless steel quality high-grade stainless steel b.s 304s31|aisi 304 7 vacuum pump · compatible with all microarray robot dry stations · ideal for pcr purification and other microplate-based kits · can be used with any vacuum manifold · excellent all-purpose vacuum pump for laboratory applications · oil-free and fully cleanroom compatible · rugged design ensures years of maintenance-free performance · in-line filter traps particles and vapors · quick release tubing connector simplifies use · uses convenient 5/16 and 3/8 8 and 10 mm vacuum tubing · vacuum meter on front of unit allows easy reading · easy access on/off switch · 110-volt and 220-volt versions for world-wide usage technical specifications · 0.8 bar vacuum pressure maximum vacuum 440mbar-550mbar · 2.1 cubic feet per minute 70 liters/min air flow · 45 decibel db noise level at 1 meter · input hose fitting ports of 10mm 6mm or 3/8 1/4 · 2 meter 6 ft power cord · weight 15 lbs 7 kg · size l w h 12.4 x 7.1 x 8.7 in 315 x 180 x 220 mm · ce approved · low power consumption of 95 watts · u.s version cat avp 110 volts at 60 hz · overseas version cat avp220 220 volts at 50 hz 8 antibiotic disc dispenser 06 hole 1 3
[close]
p. 7
9 hydrometer 10 incubator · microbiological incubator to have capacity of approx 100 litres · forced air circulation by means of a fan · work chamber to be stainless steel and to have an inner glass door · chamber to have radiused corners · outer case to be of rust proof material and finished with epoxy coating · to allow time/temperature profiles to be programmed within one cycle · operating temperature variable +4ºc to 70ºc ± 0.3ºc to be provided with 2 shelves 1 2 11 hot air oven with radial air circulation and auto shut facility size of working chamber 12 inch x 12 inch x 12 inch material of construction m.s temperature range ambient to 250ºc air circulation radial fan it saves power at least 20 by recirculating the same hot air inside the oven timer digital presettable with range upto 999 minutes paint powder coated power supply 230 volts 50hz single phase 12 seitz filter sheets · low beta glucan containing sheets · low non-specific protein binding · high dirt hold capacity · good membrane protection · long life · well graded sheets over the entire range 13 colony counter · electronic hand held colony counter · to provide an accurate count shown on clear lcd display · to count up to 9999 · to operate on single n type battery · to have an audible indicator · to be supplied in an abs case 14 digital balance balance electronic 400g cap · the balance to weigh up to 400g in 1mg increments 1 2 1 1
[close]
p. 8
· the balance to be small flat and conveniently housed in a portable lightweight easy to clean housing · to have a 7 digit luminous display with overload protection · to come complete with data input and output terminals and a print out terminal part 10 equipment instruments required in chemical pathology diagnostic laboratory 1 vortex mixture 1 · vortex mixer to operate when pressure is applied to the rubber cup which receives the vessel containing the materials to be mixed · variable speed control to accept vessels up to 25mm diameter 2 ph meter electronics · unit to have buffer sachets electrode arm electrode lead and shorting plug · available modes · ph mv absolute relative mv temperature timer 4 digit display 8 character alphanumeric prompting led display measurement of ph millivolts and temperature digital balance balance electronic 400g cap · the balance to weigh up to 400g in 1mg increments · the balance to be small flat and conveniently housed in a portable lightweight easy to clean housing · to have a 7 digit luminous display with overload protection · to come complete with data input and output terminals and a print out terminal water bath 100c 4 and 8-hole models enlarge from 7/8 inch to maximum opening diameters of 5 and 6 inches respectively · copper-clad immersion heaters provide bath temperatures up to 100°c · easy to operate with a 4-position switch and 3 different heat settings · adjustable water level regulator maintains bath depth at the desired level · stainless-steel cabinets with flanged cabinet top for table-top installations · rugged epoxy-coated corrosion-resistant stand · accommodates most laboratory glassware 1 3 1 4 1
[close]
p. 9
5 blood mixer description of function the system is used to collect donated blood from the donor at the same time mixing the blood for quality collection of blood operational requirements it is meant for stationary and mobile use gentle mixing and control of collection time to give high quality blood platelets suitable for all blood bags on the market automatic check on blood flow and collection time with buzzer alarm shall continuously display collected volume flow and time during collection shall provide repetitive notification of completed collection every minute including gentle mixing to avoid coagulation 2.2 should be technologically advanced with modern blood collection scale and mixer .it should be light weight portable technical specifications volume setting pre-selection of volume to be collected tarring of bag volume before collection tarring range 0 600 g automatic storage and recall of set volume measure volume with best accuracy indications and alarms · led indication on commencement of collection · led indication and audible alarm at the end of collection · indication of time taken for collection · indication of blood flow with audio alarm when blood flow is higher or lower than desired continuous display of collected volume flow and time during collection automatic clamping at termination of preset volume collection automatic release of bag when lifted continuous agitation of blood bags during collection 12 16 rpm easy provision to change preset volume should operate on mains as well as rechargeable battery on battery it should operate for a minimum of 5-8 hours pause facility to pause the collection optional memory to store donation data such as donor number name of operator volume collected should measure volume with accuracy less than 1 system configuration accessories spares and consumables blood mixer collection unit-01 environmental factors the unit shall be capable of operating continuously in ambient temperature of 10 400 c and relative humidity of 15-90 the unit shall be capable of being stored continuously in ambient temperature of 0 400 c and relative humidity of 15-90 shall meet iec-60601-1-2 :2001or equivalent bis general requirements of safety for electromagnetic compatibility power supply power input to be 220-240v ac 50hz as appropriate fitted with indian plug 2
[close]
p. 10
resettable overcurrent breaker shall be fitted for protection suitable automatic voltage regulator/stabilizer meeting isi specifications should be supplied broad specifications are automatic type input 150-280v output 220 v 7 50 hz single phase ac with automatic 2-4 sec cut off and 6-9 minutes restart delay quick start arrangements for bypassing the start delay suitable mcb on input voltmeter and indicators on front panel input power cable with 15 a plug and six way output terminal strip for two outlets part 11 equipment instruments required for histopathology diagnostic laboratory 1 vortex mixture · vortex mixer to operate when pressure is applied to the rubber cup which receives the vessel containing the materials to be mixed · variable speed control to accept vessels up to 25mm diameter ph meter electronics · unit to have buffer sachets electrode arm electrode lead and shorting plug · available modes · ph mv absolute relative mv temperature timer 4 digit display 8 character alphanumeric prompting led display measurement of ph millivolts and temperature hydrometer tissue floating water bath bath inside black teflon-coated heating capacity approx 65 w bath with loose stainless steel lid heating by means of heated foil double over temperature security integrated in heated circuit bath fitted in outside wall outside wall powder coated electronic temperature regulator with digital temperature display fitted flush in the front wall cee-supply socket with main switch and fuses mounted at the rear temperature selection up to 100°c comfortable user friendly spout technical data bath contents approx 3,5 litres weight and dimensions dimensions in mm dimensions bath in mm dimensions packed in mm 2 2 1 3 4 1 2
[close]
p. 11
depth x width x height 415 x 275 x 90 325 x 265 x 65 500 x 400 x 200 weight net weight packed 3 kg 6 kg electrical parameters art no 13000200 230 v 50 hz art no 13000201 230 v 60 hz art no 13000202 115 v 60 hz art no 13000203 100 v 50 hz art no 13000204 100 v 60 hz 5 microscope · standard microscope · binocular head adjustable inter pupillary distance · rotating quadruple nose piece · adjustable stage with vernier scale and spring clip for slides · co axial controls · condenser carrier vertically adjustable · brightfield condenser n.a 0.9 flip top · aperture iris diaphragm · kohler illumination · halogen illuminator 6v 20w eyepiece lenses:· x 8 widefield focusable · x 10 widefield focusable objective lenses:· x 10 achromat n.a 0.25 · x 40 achromat n.a 0.65 · x 100 achromat n.a 1.25 oil immersion options and accessories:· blue iris aperture filter · oiler · spare bulbs x 10 · carrying box · dust cover moist chamber the moist air humidifying chamber is designed for the infusion of humidified air during phakic fluid air exchange sterile single-use 10 per box most popular size or model fna plunger aluminum u-channel · 1 x 1 x 2 aluminum block 10 6 6 7 1
[close]
p. 12
8 9 · 1/8 aluminum sheet · 0.156 aluminum sheet · 1/8 stainless steel rod · 3.58º r.h 1 lead/0.333 p.d worm · 1.042 p.d worm gear · wheel gear with tapped holes · #8-32 machine screws · #8 washers · #8-32 hex nuts · #8-32 lock nuts · #8-32 electrical ring terminals · 1/8 nylon bearings · 8-19v dc motor · 10ml syringe · 12 luer tubing extension · 3-way luer stopcock · epoxy · power supply coated slide and cyto funnel for cell spin double head microscope consisting of · standard microscope · binocular head adjustable inter pupillary distance · rotating quadruple nose piece · adjustable stage with vernier scale and spring clip for slides · co axial controls · condenser carrier vertically adjustable · brightfield condenser n.a 0.9 flip top · aperture iris diaphragm · kohler illumination · halogen illuminator 6v 20w eyepiece lenses:· x 8 widefield focusable · x 10 widefield focusable objective lenses:· x 10 achromat n.a 0.25 · x 40 achromat n.a 0.65 · x 100 achromat n.a 1.25 oil immersion options and accessories:· blue iris aperture filter · oiler · spare bulbs x 10 · carrying box · dust cover 1 1
[close]
p. 13
10 object marker any which is compatible with the microscope 11 slide trays · immerse 20 slides in liquid during blocking denaturating and post-arraying · it separates and prevents slide damage during shipping and processing · optimally designed rack spacing for easy slide removal · fits standard microtiter plate formats for centrifuging · manufactured to prevent slide contamination · a convenient removable handle available 12 mold for embedding 13 staining basket 14 staining set · slide staining trough glass with lid · slide staining rack to hold 25 slides of 76x26mm size 15 fume hood laboratory fume hoods manufactured by kewaunee scientific corporation mott manufacturing labconco or fisher hamilton meeting the requirements listed below are strongly preferred hoods from other manufactures may be used if they meet the standards below and are approved by ehrs adequate documentation must be provided to ehrs in order to approve the hoods materials laboratory fume hoods shall be constructed of non-combustible nonporous material selected to resist corrosion for the service intended design and construction features the following design features from ansi/aiha z9.5 are required constant volume bypass hood no larger than 6 foot in length unless vav is specified equipped with an alnor model 405 fume hood flow alarm the work surface of the bench is recessed 1/2-inch or more below the front edge of the bench and the sides and back of the bench are provided with a seamless vertical lip 1/2-inch high an airfoil with a 90-degree arc with a minimum radius of 1-1/2 inches 1-inch air gap between the foil and the corner of the bench shall be provided at the front edge of the bench external to the sash the side posts of the front face are provided with an airfoil except that side posts wider than 3-1/2 inches are substituted for a 45-degree taper at least 4 inches long at the forward inside edge of the face utilities serving the hood shall have valves and switches at readily accessible locations outside the hood additional valves and switches also 3 5 1 2 2 1
[close]
p. 14
may be located inside the hood all valves shall meet asme or jic codes and all switches should meet nec codes the rear and top interior of the hood should be furnished with baffles to provide at least two preferably three slots adjustable in width to attain reasonably uniform face velocity under different conditions of hood use as related to heat sources size and configuration of equipment within the hood and other factors there must be screen at the bottom baffle to prevent small papers from entering the ductwork screens will have a oneinch grid size connections for exhaust ducts should be provided in the walls or roof of the hood behind the adjustable baffles the connectionsshould have a flow cross-section sized for a velocity of 2000 fpm maximum at the rated volume flow of the hood with the front sash wide open and should be located so the adjustable baffles are effective in attaining uniform face velocity the open face of the hood should be provided with one vertically movable sash capable of closing the entire front face constructed of shatterproof flame-resistant material a sash lock swivel lock that can be disengaged is required at the maximum operating height of 18 inches the vertically moving sash should be movable throughout its travel by application of no more than 5 pounds of force and should remain stationary when force is removed the fume hood shall be tested prior to leaving the factory and meet the ashrae 110-most current tracer gas test protocol with a rating of 4 am 0.05 documentation shall be provided the hood shall be equipped with an airflow monitor that measures flow and has both visual and auditory alarm the alarm should be provided and installed by the fume hood manufacturer installation hoods shall be tested after installation and meet ashrae 110-most current tracer gas test protocol with a rating of 4 ai 0.1 and the flow vizualization and velocity procedure tests a professional engineer shall seal test reports reports shall be provided to ehrs the average face velocity at 18 inches shall be 100 fpm plus or minus 10 fpm with low flow alarms set at 80 fpm no individual face velocity measurement shall be more than plus or minus 20 of the average flow of the hood nor shall any one measurement point be less than 80 or higher than 120 fpm on off switches for blowers are not permitted directional supply air diffusers shall not discharge air toward the hood nor shall they be installed within 10 feet from the hood casework under hood a factory mutual approved ul listed flammable liquids storage cabinet in compliance with nfpa 30 and osha regulations shall be installed under the fume hood this cabinet shall be 48 inches in width the cabinet shall be vented to the hood plenum the shall be equipped with a flash arrester labeling the hood shall be labeled with the exhaust fan number and location.
[close]
p. 15
part 12 equipment instruments required for hematology diagnostic laboratory 1 centrifuge 3 · angle rotor · maximum speed 6000rpm rcf 6250g · timing up to 99 hours continuous · programmable by keypad 32 programme storage · digital display of programme speed timing braking · imbalance detection · lid safety interlock · angle rotor maximum capacity 24 x 15ml 24 x 7ml with adaptor 1 part 13 equipment instruments required in environment research laboratory automatic absorption spectrometer 1 flame only · graphite furnace · combined flame and graphite furnace with the option for up to nine heating ramps the system is fully controlled from a pc with aawin software the main features of the system are · the integration of flame burner and graphite furnace in one compact unit enables automatic interchanging between the flame mode and furnace mode using few simple keystrokes · eight lamp turret for automatic positioning of each hollow cathode lamp · automatic fuel gas c2h2 flow rate adjustment to optimize combustion in element analysis · automatic adjustment of burner height for optimized analysis conditions · automated wavelength scanning and peak searching · automatic reflecting/transmittance beam-splitter positioning · automatic switchover of spectral bandwidth 5 choices · automatic pmt and lamp current adjustment automatic aa and bg energy balance wavelength range 190nm 900nm monochromator czerny-turner configuration spectral bandwidth 0.1nm 0.2nm 0.4nm 1.0nm 2.0nm 5 steps with automatic changeover wavelength accuracy ±0.25nm wavelength repeatability 0.15nm baseline stability 0.005a/30 mi to be provided with the following additional facilities 1 hydride generator,